AED Plus ZAS Software Download


AED Plus® ZOLL® Administrative Software (ZAS) can be used to update the software on your AED Plus to the latest revision or to adjust the default configuration.

AED Plus Software version 6.24 or higher fully complies with the latest American Heart Association (AHA) and European Resuscitation Council (ERC) Guidelines.

If your AED Plus is running v6.24 or higher, you don't need to do anything. Any AED Plus, no matter how old, can be upgraded to v6.24. For information about how to order complete upgrade services, contact your ZOLL Sales Representative or your Authorized ZOLL Distributor.

How to determine the software version

AED Plus units manufactured after April 2011, when SW v6.24 was released, are fully compliant with the latest AHA and ERC Guidelines.

You can determine the manufacturing date of your AED Plus by referring to the two numbers and letter following the "X" in the serial number. The two numbers following the X indicate the year of manufacture, and the letter following the year indicates month of manufacture. For example, serial number X12B123456 would have been manufactured in February of 2012.

What it takes

You need to have ZOLL Administrative Software (ZAS) v6.24 installed on your PC, and you will need an infrared port (IrDA device) to communicate with your AED Plus. With ZAS and an IrDA port, you can easily upgrade the software in any AED Plus to v6.24 or adjust the device's configuration if needed.

If you are updating the AED Plus software for the ERC 2015 Guidelines that's all you need to do — upgrade the software.

If you are updating the software for the AHA 2015 Guidelines, after upgrading your AED Plus software, you may want to apply two peel-off-stick-on labels to the faceplate of your AED Plus to cover up the two icons that are no longer used in the AHA protocol. Note: Applying these labels is optional. Your AED Plus will perform perfectly without these labels.

If you apply the labels, the face of your AHA 2015 Guidelines-compliant AED Plus will look like this:

AED Plus cover upgrade

To get started:

Download v6.24 Upgrade Software (will require a valid serial number from the AED Plus), purchase stick-on labels, and/or an IrDA infrared port.

After upgrading your AED Plus:

You must register your upgrade with ZOLL to preserve your warranty and make certain you receive notification of any Device Corrective Actions that may occur in the future. Registration data is NOT forwarded to any other organization under any circumstances, and is NOT used for marketing purposes without your permission.